SKU
HI99300

Prenosni merilnik prevodnosti - HI99300

385,00 €
Na zalogi

Designed to bring simplicity to testing in manufacturing and environmental industries, the HI99300 Low Range EC/TDS Meter is designed for measurements in pure and low ionic strength water applications. The specialized HI763063 probe features EC/TDS and temperature in one convenient, rugged probe for low range EC measurements from 0 to 3999 µS/cm and TDS measurements from 0 to 2000 ppm.

  • ±2% EC/TDS Accuracy
  • Comes with all the necessary solutions and batteries- everything you need to get started measuring right away.
  • Perfect for water quality testing in manufacturing and environmental industries.

Low Battery Indicator

The included batteries provide 500 hours of continuous use. When the battery level is below 10%, a warning symbol will blink to indicate low battery condition. When low battery power may affect your results, the meter will shut down to avoid erroneous readings caused by low battery level.

Automatic Shut-off

The auto-off feature can be set to shut the meter off at 8 or 60 minutes, or it can be disabled so that the meter continuously stays on. The auto-off feature prolongs battery life for peace-of-mind and energy savings if the meter is accidentally left on.

Two Measurement Modes

HI99300 offers two measurement modes: EC and TDS. Users can easily switch between measurement modes with a simple press of the MODE button. EC measurements range from 0 to 3999 µS/cm; TDS measurements range from 0 to 2000 ppm.

 

 

Built to be Durable

Designed to withstand the knocks, drops, and spills of real life, the new IP67 body ensures top performance in any environment. These meters are protected against dust and water intrusion from any direction.

Portable Design

Made to fit comfortably in your hand for on-the-go testing, the multilevel LCD display provides at-a-glance readings of EC/TDS and temperature from any angle and dual button operation streamlines the testing process for any user.

Quick Connect Probe

A Quick Connect DIN connector makes attaching and removing the probe simple and easy. The rubber coating protects the cable and provides a water-tight seal without twisting.

 

Your measurements require detailed attention; so should your electrode. The HI99300 Low Range EC/TDS Meter features a replaceable HI763063 probe that has been carefully designed for use in applications as diverse as wastewater treatment, reverse osmosis, and HVAC.

Two-Ring Probe

The HI763063 is a two-ring probe that comes with the HI99300, offering an accurate solution for low range EC and TDS readings.

Built-in Temperature Probe

An integrated temperature probe ensures fast, accurately compensated readings even during sudden temperature fluctuations.

 

 

We believe that buying a meter should be simple. Your Low Range EC/TDS Meter is equipped with everything you need for successful testing right out of the box.

1413 μS/cm and 1382 ppm Conductivity Standard (1 each)

When performing an EC/TDS calibration it is important that the correct standards are used. We include some starter standards to get you going. The single-use sachets are individually sealed so you use fresh buffer each time you calibrate.

 

HI99300 Models and Compatible Accessories

Old HI99300 (Left)

  • Uses the HI76306D EC/TDS probe with DIN Connector
  • Compatible with the HI710023 Shockproof Boot

New HI99300 (Right)

  • Uses the HI763063 EC/TDS probe with Quick DIN Connector
  • Compatible with the HI710028 Shockproof Boot

*The new Quick DIN probe connection is compatible with the old body design, however, the old DIN probe connections are not compatible with the new body design.

 

 

Specifications

Specification Name Detail
sku HI99300
Temperature Range -5.0 to 105.0 °C / 23.0 to 221.0 °F
Temperature Resolution 0.1 °C / 0.1 °F
Temperature Accuracy (@25°C/77°F) ±0.5 °C / ±1.0 °F Temperature
EC Range 0 to 3999 μS/ cm*
EC Resolution 1 μS/cm
EC Accuracy (@25°C/77°F) ±2% F.S. (EC/TDS)
EC Calibration Automatic, one-point at 1413 μS/cm
TDS Range 0 to 2000 ppm
TDS Resolution 1 ppm (mg/L)
TDS Accuracy (@25°C/77°F) ±2% F.S. (EC/TDS)
TDS Conversion Factor Selectable from 0.45 to 1.00 with 0.01 increments
TDS Calibration Automatic, at one-point at 1382 ppm (CONV=0.5) or 1500 ppm (CONV=0.7)
Temperature Compensation Automatic, with β selectable from 0.0 to 2.4 %/ °C with 0.1 increments
Power Supply 1.5V AAA batteries (3 pcs.)
Power Consumption approx. 500 hours of continuous use
Environment 0 to 50 °C (32 to 122 °F) RH max. 100%
Dimensions 155 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight 197 g (6.91 oz.) with batteries
Ordering Information Each meter is supplied complete with: • HI763063 pH/EC/TDS probe with built‑in temperature sensor, DIN connector and 1m (3.3’) cable • HI70031 1413 μS/cm and HI70032 1382 ppm 1 sachet each for HI99300 • 100 mL beaker (1 pcs.) • Alkaline batteries: 1.5V AAA (3 pcs.) • Rugged carrying case • Calibration certificate of meter • Calibration certificate of probe • Instruction manual
Warranty 2 years
Recommended Probe HI763063 EC/TDS/temperature sensor, DIN connector and 1 m (3.3’) cable (included)
Vse cene so brez DDV in brez stroškov dostave, razen če ni to kako drugače določeno. Hanna Service d.o.o. — Sermin 75H, Bertoki — 6000 Koper (Slovenjia) — davčna številka: SI92346669 © 2024 Vse pravice pridržane